Determination of the dead time of a GM counter using the two-source method

Quyet Dang Pham 1, * and Anh Thi Tu Trinh 2

1 Training Center, Nuclear Research Institute, 01 Nguyen Tu Luc, Dalat, Vietnam.
2 Office of National Assembly Delegations and People Councils, 02 Tran Hung Dao, Dalat, Vietnam.
 
Research Article
International Journal of Science and Research Archive, 2024, 13(01), 790–795.
Article DOI: 10.30574/ijsra.2024.13.1.1755
Publication history: 
Received on 09 August 2024; revised on 18 September 2024; accepted on 20 September 2024
 
Abstract: 
Calculations and experiments of the dead time and standard error of the ST360 counter have been carried out at the Training Center of Dalat Nuclear Research Institute employing the Sr-90 and Cs-137. The average dead time and standard error of the counter for measurements lasting 300 seconds or longer is (141.5 ± 11.8) μs. This value is in good agreement with that mentioned by the manufacturer.
 
Keywords: 
GM; ST360 counter; Dead time; Two source method
 